wires, or signs of burning. Replace where required
Using the stopwatch and two measured marks on the
d. Check mechanical security of all connectors,
chart insure that the chart will move exactly 71?
connector brackets, and cable clamps. Tighten where
inches in 15 minutes of a powerline frequency of 60
required.
Hz. (Set multispeed recorders to inch per minute.)
If the chart speed varies more than 5 percent, send
the unit to direct support maintenance for chart
drive motor replacement (para 4-10).
Clean the control panel interior using a vacuum
cleaner and a soft brush. Use the soft brush to loosen
any foreign matter around the contacts of switch
Check the condition of the slide-wire contacts. If ei-
S202 and plugs P201 and P202; then remove the
ther is worn halfway through, replace and adjust
loosened material with the vacuum cleaner.
Converter Interior
a. Unlatch all components and remove them from
a. Clean the interior of the signal data converter
the cabinet.
using a vacuum cleaner. Remove any deposits using
b. Clean the interior of the cabinet with a vacuum
a lint free cloth and isopropyl alcohol.
cleaner.
c. Inspect all cables and connector for frayed
Change 3
|