| TM 11-6660-204-25
the frequency standard output to test jack J303
5-6. Pen-Lifter Test
on the signal data converter. The frequency stand-
a. Test Equipment and Materials. F r e q u e n c y
ard output should be at zero. On multispeed recor-
Standard TS-65C/FMQ-1 is required for this
ders, select inch per second chart speed.
test.
c. Procedure.
b. Test Connections and Conditions. C o n n e c t
a. Test Equipment and Materials. F r e q u e n c y
Standard TS-65C/FMQ-1 is required for this
The recorded input signal to the radiosonde recor-
test.
d e r at any frequency between 10 and 190 Hz
b. Test Connection and Conditions. C o n n e c t
should not deviate from the recording of the same
the frequency standard output to test jack J303
s i g n e d by more than 0.1 of one chart division
on the signal data converter. Adjust sensitivity of
when approached by a higher or lower frequency.
the recorder as described in paragraph 3-34.
This deviation is called hysteresis. Excessive hys-
teresis is eliminated as explained in paragraph
c. Procedure.
4-12.
Change 3
|