|
| e. Apply this if. test signal to the test
set.
f. Increase the signal generator level
c o n t r o l setting to produce a full-scale
meter indication. The ratio of the two sig-
nal generator output level control settings
in db (in this procedure and that in c above)
m u s t meet the requirements of the fol-
lowing chart:
e. A f t e r d e t e r m i n i n g t h e i m a g e f r e -
quency for the frequency under test, apply
this frequency from the tuning unit signal
generator to the test set.
f. Increase the tuning unit signal gener-
g. R e p e a t t h e p r o c e d u r e s g i v e n i n c
ator level control setting to produce a full-
through f above at the middle and highest
scale panel meter indication. The ratio of
frequency points of each band for all tuning
the two signal generator output level con-
units.
trol settings in db (in this procedure and
t h a t in c above) must meet the require-
109. Rf Intermodulation
ments of the following chart:
The rf intermodulation test concerns the
r e s p o n s e of the test set to the intermod-
ulation components that are generated
a h e a d of the if. amplifier in each tuning
u n i t . The response is measured with the
test set receiving high-intensity broadband
i n t e r f e r e n c e signals from the self-con-
tained impulse generator. No signal gener-
ator is required. Make the test as follows:
a. Rotate the SIGNAL ATTENUATOR DB
g. R e p e a t t h e p r o c e d u r e s g i v e n i n c
control to the 0 CW ONLY position.
through f above at the middle and highest
b. S e t t h e c a l i b r a t i o n s w i t c h t o t h e
frequency points of each band for all tuning
SHUNT CAL position.
units.
c. Set the function switch to the PULSE
PEAK position.
108. If. Rejection
d. Set the TUNING control and the fre-
a. With a tuning unit plugged into the
quency dial of the tuning unit under test to
tuning unit compartment, connect the ap-
t h e lowest frequency of the lowest band.
e. With the IMPULSE GENERATOR DB
p l i c a b l e signal generator for the tuning
u n i t under test as directed in paragraph
A B O V E 1 V/MC switch at its OFF po-
sition, vary the tuning unit GAIN control
to produce a zero db indication on the upper
b. Set the AN/URM-85 front-panel con-
scale of the panel meter.
trols as directed in paragraph 105b.
f. Set the IMPULSE GENERATOR DB
A B O V E 1V/MC at ON and adjust the
i m p u l s e generator coarse and fine output
d. Retune the signal generator to the
c o n t r o l s to produce a 2-db deflection on
i n t e r m e d i a t e frequency of each band, as
the panel meter. Record the settings of the
listed in the chart in paragraph 107d.
224
|
Privacy Statement - Press Release - Copyright Information. - Contact Us |