Click here to make tpub.com your Home Page

Page Title: Calibration as Two-Terminal RF Voltmeter
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home

   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 



impulse g e n e r a t o r output level controls;
of the band under test; then check at two
t h e sum o f the settings is the reference
other equally spaced frequency points
--
within the band.
sensitivity.
k. Repeat the procedures in e through
g. Now disable the local oscillator in the
j for each band of the tuning unit under test.
tuning unit under test. Each tuning unit re-
quires a different procedure for this oper-
110. Calibration as Two-Terminal RF
a t i o n . Perform the procedure outlined in
the following chart:
This procedure is used to calibrate. the
t e s t set as a two-terminal rf voltmeter.
S e t up the equipment as shown in figure
94; however, use the 10-db pad instead of
the 5-db variable step attenuator. The 10-
d b pad can be connected directly to the
tuning unit signal generator.
a. If tuning units 1, 2, or 3, are used,
c o n n e c t the applicable tuning unit signal
g e n e r a t o r (recently calibrated for power
o u t p u t ) to the test set and proceed as
follows:
(1) S e t t h e f u n c t i o n s w i t c h t o C W
A V E R A G E a n d the calibration
switch to SERIES CAL &OP-
ERATE.
(2) I f tuning unit 1 is used, set the
h. With the local oscillator of the tuning
SIGNAL ATTENUATOR switch to
u n i t disabled (g above), readjust the im-
0 CW ONLY; if tuning units 2 or
p u l s e generator coarse and fine output
3 are used set the SIGNAL AT-
l e v e l control settings to again produce a
TENUATOR switch to 20 DB.
2-db indication on the panel meter. Record
(3) Set the tuning unit signal generator
t h e sum of the two output level control
output level to produce a 10-
s e t t i n g s : t h i s is the rf intermodulation
m i c r o v o l t signal at the SIGNAL
sensitivity.
INPUT terminal of the test set for
i. S u b t r a c t t h e v a l u e o f t h e r f i n t e r -
t u n i n g unit 1, or a 100 microvolt
modulation sensitivity (h above) from the
signal for tuning units 2 or 3.
r e f e r e n c e sensitivity (f above) to deter-
N o t e : If the 10-db pad is too large to
mine the difference in db. This difference
p r o v i d e a signal of proper amplitude at
in settings for the impulse generator out-
the test set SIGNAL INPUT terminal, use
p u t level controls must be the same or
a Pad of a smaller value (AT-50-8) to pro-
duce the proper signal amplitude.
h i g h e r than the values shown in the fol-
(4) Adjust the GAIN control to obtain
lowing chart:
a 20-db reading on the test set
p a n e l meter.
(5) S e t t h e f u n c t i o n s w i t c h t o C W
P E A K and the calibration switch
to SHUNT CAL.
(6) Set the IMPULSE GENERATOR DB
ABOVE lV/MC switch to ON
a n d adjust the impulse generator
output controls (coarse and fine) to
again obtain a 20-db reading on the
j. R e p e a t t h e p r o c e d u r e g i v e n i n e
t e s t set panel meter.
through i above for the high-frequency end
(7) Record the settings of the impulse
225


Privacy Statement - Press Release - Copyright Information. - Contact Us

Integrated Publishing, Inc. - A (SDVOSB) Service Disabled Veteran Owned Small Business