| TM 11-6625-924-14
CHAPTER 3
OPERATING INSTRUCTIONS
3-1. Controls, Indicators, and Connectors
3 - 1 . Controls, Indicators, and Connectors
(fig. FO-1) (cont)
Item
Function
Item
Function
flag or annunciator circuit in
Synchro transmitter No. 1 con.
Controls rotation of pointer No. 1
indicator.
trol
and pointer No. 2 on indicator
Connects output of B data
B307-B308-B304 switch
under test to check their opera-
transmitter B307, A data
tion and calibration.
transmitter B308, or heading
Meter
Used during flag and annunciator
selector transmitter B304 of in-
tests of the indicator to indicate
dicator to adjacent X-Y-Z bind-
current or a null.
ing posts to allow testing of
Synchro transmitter No, 2 con-
Controls rotation of slaving con-
transmitter output (POINTER
trol
trol transmitter in indicator
NO. 1 & 2-ALL OTHER TESTS
under test to check its opera-
switch in ALL OTHER TESTS
tion and tuning.
OHMS control
position).
Indicates resistance when
POINTER ADF-VOR-RMI
Connects synchro transmitter No.
BRIDGE SEL switch is in
switch
1 output signal to ADF, VOR,
FLAG RES or ANNU RES posi-
or RMI input of pointer No. 1
tion.
PRESS-TO-READ switch
on the indicator under test
When pressed, allows meter to in-
(POINTER NO. l-NO. 2-B306
dicate.
CUR ADJ control
switch in No. 1 position). Also
Adjusts current through flag and
connects 26 VAC to pointer No.
annunciator circuits in indica-
1 (POINTER NO. 1 & 2-ALL
tor under test.
OTHER TESTS switch at
POWER indicator
Indicates power is applied to test
POINTER NO. 1 & 2 position).
panel (ON-OFF switch at ON
POINTER NO. l-NO. 2-B306
position).
switch
ON-OFF switch
transmitter No. 1 output signal
Applies power to test panel,
HIGH POT binding posts
to pointer No. 1 of indicator
Provides connection for ohm-
under test (through ADF-VOR-
meter during insulation break-
RMI switch). Also connects 26
down (HIGH POT) test of in-
VAC to input of synchro
dicator (HIGH POT-NORM
transmitter No. 1.
switch in HIGH POT position).
HIGH POT-NORM switch
HIGH POT position. Connects in-
transmitter No. 1 output signal
dicator to HIGH POT binding
to pointer No. 2 of indicator
posts for high pot test.
under test. Also connects 26
NORM position Connects indica-
VAC to input of synchro
tor to test circuits in test panel
transmitter No. 1.
during all other tests.
PIN SEL switch
B306 position. Connects synchro
Connects selected indicator con-
transmitter No. 1 output signal
nnector pin(s) to HIGH POT
to slaving control transmitter
binding posts during high pot
B306 in the indicator under test
test (HIGH POT-NORM switch
to simulate flux gate input sig-
in HIGH POT position).
POINTERS 1 & 2-ALL
nals. Also connects 1.5 VAC to
1 & 2 position. Connects 26 VAC
OTHER TESTS switch
input of synchro transmitter
to pointer No. 1 and No. 2 for
No. 1.
pointer tests of indicator.
1.5 V ADJ control
Adjust 1.5 VAC Supply of test
ALL OTHER TESTS position.
panel.
Connects 26 VAC to B data
1.5 V binding posts
Used during adjustment of 1.5
transmitter B307, A data
VAC supply of test panel.
5 VDC binding posts
Provides connection-for external 5
selector transmitter B304 in in-
VDC input for checking lights
dicator during tests of those
in indicator.
transmitter.
B306
X-Y-Z (B307-B308-B304) Connect to B data transmitter
X-Y-Z binding posts
binding posts
Used to apply normal load (flux
B307, A data transmitter B308
valve) to slaving control
or heading selector transmitter
transmitter B306 in indicator
B304 (through B307-B308-B304
during testing.
switch) to allow individual test-
VTVM binding posts
Used to apply signal to slaving
ing of transmitter outputs.
control transmitter B306 dur-
BRIDGE SEL switch
Connects bridge circuit for cur- ,
ing testing.
rent or resistance readings of'
3-1
|