| TM 11-5820-695-35
Step
Control Setting
No Test Equipment
Equipment Under Test
Test Procedure
Performance Standard
1 SIGNAL GENERATOR
Converter-Multiplier
HP 626A
CV-3633/GRC-144(V)
MODE SELECTOR:
BAND SET switch to
CW
position 1
SYNC SELECTOR:
Bandpass filter
INT FM
3A3:14.4 GHz
OUTPUT ATTEN
3A4:14.8 GHz
DBM: 0.0
FREQ DIAL: 14.4
GHz
PWR: CCW
FREQUENCY COUNTER
HP 5340A
RANGE: 250 MHz -
18 GHz
RESOLUTION: 10 KHz
POWER METER HP 432A
MOUNT RESISTANCE:
100 ohms
CALIBRATION FACTOR:
Calibrate same as in
step 1
RANGE 0 DBM
a. Connect the HP
a. N/A.
a. N/A.
a. None.
486A thermistor
NOTE
mount to the di-
Do not adjust signal-
rectional coupler
l generator out-
coupled output.
put level during
Adjust signal
performance of this
generator FREQ
test.
DIAL for 14.4
NOTE
GHz on frequency
Thermistor Mount
counter and the
Hewlett-Packard
PWR SET DIAL for
HP 478A is used to
-20.0 DBM on
connect frequency
power meter.
converter 3A2 to
power meter.
b. Increase sig-
b. N/A.
b. Disconnect power b. None.
nal generator
meter from direct-
OUTPUT ATTEN
ional coupler and
DBM dial until
connect power meter
it reads -10
to 3A2AT2J2 RF OUT
DB. (-30 DBM
connector on elec-
input to con-
tronic frequency
verter).
converter 3A2 as
shown in figure 3-40.
Change 6
3-114
|