TM 11-5820-695-35
(para 3-12b, item No.  81) for corrective action upon
o.
Remove the protective plate over the
completion of this procedure.
potentiometer  bracket,  and  adjust  potentiometer
2A15A2R5 (fig. 3-18) for a 2A15A2M1 meter indication
k.
Turn the TS-656/U off and remove the
of 30. If a meter indication cannot be obtained, refer to
connector adapter from connector 2A16WlXA12A.
the troubleshooting chart (para 6-12b, item No. 3) for
I.
Reinstall alarm monitor 2A12 on plate
corrective action. Do not continue this procedure.
assembly 2A16 (TM 11-820-695-12).
p.
Disconnect the test equipment from coaxial
m.
Set POWER ON/OFF switch 2A15A2S1 on
cable 2W9.
meter panel assembly 2A15A2S1 to ON.
q.
Connect  coaxial  cable
2W9
to
local
3-65. Calibration of R-1467(P)/GRC-144(V) and R-
oscillator bandpass filter 2FL6.
1467(P)A/GRC-1440V)  Local  Oscillator  Metering
r.
Turn the AN/URM-2B off.
Circuits
NOTE
a.
Turn Test Set, Radio Frequency Power AN/
Do not calibrate any of the following
USM-161 and Signal Generator AN/URM-2B on and
R-1467(P)/GRC-144(V)local oscillator
allow a 30-minute warmup period.
metering circuits if the LOCAL OSC
b.
After the warmup period, calibrate the
POWER metering circuit (step o
AN/USM-161 using the procedures given in TM 11-
above) could not be adjusted to
6625-498-12 and calibrate the AN/ URM-52B at 4700
obtain a 2A15A2M1 meter indication
MHz using the procedures given in TM 11-6625-214-10.
of 30.
c.
Connect the equipment as shown in figure
s.
To calibrate any of the following R-1467(P)/
GRC-144(V) local oscillator metering circuits, turn meter
d.
On the AN/USM-161, set the BIAS-READ
selector switch 2A15A2S1 (fig. 3-18) to 'the circuit to be
switch to READ, turn the POWER RANGE switch to 3.0
calibrated  and  observe  the  indication  on  meter
MW +5 DBM, turn the COMP ATTENUATOR dial to 0,
2A15A2M1. If meter indication is not 100, remove the
and turn the POWER indicator dial for a dial scale
protective plate over the potentiometer bracket and
indication of 1.0 on the center scale.
adjust the associated potentiometer (listed below) for a
e.
Adjust the OUTPUT ATTEN control on the
2A15A2M1 meter indication of 100.
If a meter
AN/URM-2B for a null indication on the AN/USM-161
indication of 100 cannot be obtained, refer to the
NULL INDICATOR meter.
troubleshooting chart (para 3-12b, item No. 33 through
36) for corrective action.
f.
Adjust the POWER SET control on the
AN/URM-2B for an index indication of 0 dbm on the
Circuit to be calibrated
Potentiometer(fig. 3-4)
OUTPUT ATTEN dial.
FREQ MIXER CURRENT 1............................ 2A15A2R3
g.
Turn the AN/;USM-161 off and disconnect
FREQ MIXER CURRENT 2............................ 2A15A2R4
RF Cable CG-92D/U from Adapter UG-29B/U (fig. 5-2).
MULT-OSC ................................ ....................2A15A2R6
h.
Set the OUTPUT ATTEN control on AN/
t.
On R-1467/GRC-144, reset the controls on
URM-52B for a dial scale indication of -6 dbm.
2FL6 and 2A14 to the operating frequency recorded in
step i.
i.
Record
the
R-1467(P)/GRC-144(V)
operating frequency.
3-66. R-1467(P)/GRC-144(V) and R-1467 (P)A/GRC-
144() Carrier Alarm Adjustment
j.
On  R-1467(P)/GRC-144(V),  set  local
oscillator bandpass filter 2FL6 and electrical frequency
NOTE
synthesizer 2A14 controls to 4700 MHz.
Performing this procedure will affect
k.
Disconnect coaxial cable 2W9 (fig.  324)
the calibration of the T-1054(P)/GRC-
from local oscillator bandpass filter 2FL6, I. Connect
144(V) radio test set.
Therefore,
the equipment as shown in figure 3-27.
perform the radio test set calibration
procedure given in paragraph -61
m.
Turn meter selector switch 2A15A2S1 on
meter  panel  assembly  2A15A2  to  LOCAL  OSC
upon completion of the following
POWER.
procedure.
i.
Fine tune the SIGNAL FREQUENCY control
a.
Turn Test Set, Radio Frequency Power
on AN/URM-2B for a peak indication on meter
AN/USM-161 and Signal Generator AN/URM-52B ON
2A165A2M1.
and allow a SO-minute warmup period.
Change 6
3-81